It is plain that any temperature which is deemed most convenient might be chosen for a particular laboratory, but it cannot be too strongly emphasized that all measuring instruments, including burettes, pipettes, and flasks, should be calibrated at that temperature in order that the contents of each burette, pipette, etc., shall be comparable with that of every other instrument, thus permitting general interchange and substitution.

Fill the burettes with the standard acid and alkali solutions, noting the initial readings of the burettes and temperature of the solutions. Run in acid from the burette, stirring and avoiding loss by effervescence, until the solution has become pink.

The treatment described above will usually accomplish this, but, in the case of plain burettes it is sometimes better to allow a little of the liquid to flow out of the tip while it is bent upwards. Any air which may be entrapped then rises with the liquid and escapes. All liquids when placed in a burette form what is called a meniscus at their upper surfaces.

It is plain that the standard solutions must be scrupulously protected from concentration or dilution, after their value has been established. Accordingly, great care must be taken to thoroughly rinse out all burettes, flasks, etc., with the solutions which they are to contain, in order to remove all traces of water or other liquid which could act as a diluent.

Repeat until the ratio of the two solutions is established, remembering all necessary corrections for burettes and for temperature changes. Commercial iodine is usually not sufficiently pure to permit of its use as a standard for thiosulphate solutions or the direct preparation of a standard solution of iodine.

Two burettes, one at least of which should have a glass stopper, are required throughout the volumetric work. Both burettes should be calibrated by the student to whom they are assigned. Record the weight in the notebook. Do not attempt to stop exactly at the 10 cc. mark, but do not vary more than 0.1 cc. from it.

PROCEDURE. Fill a glass-stoppered burette with the permanganate solution, observing the usual precautions, and fill a second burette with the ferrous sulphate solution prepared for use with the potassium bichromate. The permanganate solution cannot be used in burettes with rubber tips, as a reduction takes place upon contact with the rubber.

Prescott, surrounded by his retorts, crucibles, burettes, and condensers, received us much more graciously than I had had any reason to anticipate. He was a man in the late forties, his face covered with a thick beard, and his eyes, which seemed a little weak, were helped out with glasses almost as scholarly as ours.

Burettes may also be calibrated by drawing off the liquid in successive portions through a 5 cc. pipette which has been accurately calibrated, as a substitute for weighing.

Obtain the ratio of the sodium hydroxide solution to that of hydrochloric acid by dividing the number of cubic centimeters of acid used by the number of cubic centimeters of alkali required for neutralization. If the variation in results is greater than this, refill the burettes and repeat the titration until satisfactory values are obtained. Use a new page in the notebook for each titration.